FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE

MDR report key: 3834563 · Received March 19, 2014

Report

Report Number
1419937-2014-00282
Event Type
Malfunction
Date Received
March 19, 2014
Date of Event
February 18, 2014
Report Date
February 18, 2014
Manufacturer
MEDELA, INC.
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. UPON RECEIPT OF THE PRODUCT, A BREACH WAS FOUND IN THE TRANSFORMER. THE CAUSE OF THE BREACH IN THE REV N TRANSFORMER HAS NOT BEEN DETERMINED AT THIS TIME. IT IS CURRENTLY BEING INVESTIGATED UNDER (B)(4). A VISUAL EXAMINATION OF THE POWER SUPPLY REVEALED THE TRANSFORMER HOUSING WAS CRACKED OPEN, EXPOSING INNER ELECTRICAL CIRCUITRY. A VISUAL EXAMINATION OF THE CORD REVEALED NOTHING UNUSUAL. THE POWER SUPPLY WAS PLUGGED IN AND THE VOLTAGE ACROSS THE DC PLUG WAS MEASURED AT 13.60 VDC, WHICH IS NORMAL AND INDICATES THAT THE POWER SUPPLY IS PRODUCING THE CORRECT VOLTAGE. RESISTANCE ACROSS THE DC PLUG WAS MEASURED OPEN, WHICH INDICATES THAT THERE IS NOT A SHORT IN THE DC CORD. SMOKE, FIRE AND SPARKING WERE NOT OBSERVED WHILE THE POWER SUPPLY WAS PLUGGED IN. THE RESISTANCE ACROSS THE AC BLADES MEASURED AS 57.1 OHMS, WHICH IS NORMAL AND INDICATES THAT THE THERMAL FUSE DID NOT BLOW.

Description of Event or Problem · 1

CUSTOMER REPORTED TO CUSTOMER SERVICE THAT THE PRONGS ON THE TRANSFORMER OF HER PUMP IN STYLE PUMP WERE PUSHED INTO THE HOUSING; HOWEVER, WHEN THE PRODUCT WAS RECEIVED AND EVALUATED BY QE, A BREACH IN THE TRANSFORMER WAS FOUND EXPOSING UNDERLYING ELECTRONICS WHICH IS A SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163309 PUMP IN STYLE ADVANCED BREAST PUMP GO TOTE HGX MEDELA, INC. 57063/9207010 1313/REV N