SYSTEM 7 DUAL TRIGGER ROTARY
Report
- Report Number
- 0001811755-2014-01956
- Event Type
- Malfunction
- Date Received
- May 28, 2014
- Date of Event
- May 21, 2014
- Report Date
- May 21, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS RECEIVED FOR EVALUATION; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED. THE DEVICE WAS RECEIVED FOR EVALUATION; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.
THE REPORTED RUN-ON OF THE DEVICE WAS DUPLICATED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH FUNCTIONAL EVALUATION. UPON DISASSEMBLY, IT WAS FOUND THAT THE SHAFTS OF THE TRIGGERS WERE CORRODED, WHICH CAN CAUSE THE REPORTED EVENT.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 314157 | SYSTEM 7 DUAL TRIGGER ROTARY | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT | HWE | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |