FDA Adverse Event Malfunction Summary report: N

VASO VIEW HEMOPRO

MDR report key: 3834494 · Received March 18, 2014

Report

Report Number
2242352-2014-00292
Event Type
Malfunction
Date Received
March 18, 2014
Date of Event
February 21, 2014
Report Date
February 21, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE SILICONE LAYER OF THE BALLOON HAD RUPTURED. WHILE WE ARE UNABLE TO CONCLUSIVELY DETERMINE A ROOT CAUSE, EACH BTT GOES THROUGH A FULL INFLATION/DEFLATION INSPECTION PRIOR TO DISTRIBUTION. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. A LOT HISTORY RECORD REVIEW COULD NOT BE PERFORMED AS THE PRODUCT LOT NUMBER IS UNK. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO BTT SHORT PORT BURST AFTER PLACING MINIMAL AIR. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
159659 VASO VIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000 NI

Patients

Seq Age Sex Outcome Treatment
1 NI