FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3834468 · Received March 17, 2014

Report

Report Number
1720753-2014-02393
Event Type
Malfunction
Date Received
March 17, 2014
Date of Event
February 28, 2014
Report Date
March 17, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE GIB FFB, POWER CONTROL PCB, AND SYSTEMS INTERFACE PCB ASSEMBLIES WERE EVALUATED AND REPLACED. THE ASSOCIATED POWER SUPPLY WAS ALSO REPLACED AS PART OF THE SERVICE EVENT. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM FAILED TO BOOT TO A USABLE STATE EXHIBITED A NON-RECOVERABLE LOSS OF FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156837 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1