FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3834458
·
Received March 17, 2014
Report
- Report Number
- 1720753-2014-02383
- Event Type
- Malfunction
- Date Received
- March 17, 2014
- Date of Event
- February 26, 2014
- Report Date
- March 17, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE BOARDS AND INTERCONNECT CABLE CONNECTORS WERE EVALUATED AND RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR RESULTED IN THE INABILITY TO GENERATE FLUOROSCOPIC X-RAY. THERE IS NO REPORT OF A PT INJURY OR DEATH ASSOCIATED WITH THIS EVENT.