FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3833994 · Received May 27, 2014

Report

Report Number
2955842-2014-03253
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 2, 2014
Report Date
May 2, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION WAS NOT ABLE TO REPLICATE THE COMPLAINT. THE INSTRUMENT WAS PLACED ON AN IN-HOUSE SYSTEM AND PASSED RECOGNITION AND ENGAGEMENT TEST. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIPS OPENED AND CLOSED PROPERLY. ADDITIONAL OBSERVATION NOT REPORTED WAS THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI HYSTERECTOMY PROCEDURE, THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT QUIT WORKING. ANOTHER INSTRUMENT WAS OPENED TO CONTINUE WITH THE CASE. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED. NOTHING REPORTEDLY FELL INTO THE PATIENT.