FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 3833868 · Received May 27, 2014

Report

Report Number
1416980-2014-17118
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 5, 2014
Report Date
May 2, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. ALL RETURNED HOMECHOICE (HC) DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE DEVICE PASSED RITE TESTING. NO ABNORMALITIES WERE IDENTIFIED DURING INTERNAL/EXTERNAL INSPECTIONS. THE PROBLEM OF A HIGH DRAIN ALARM WAS IDENTIFIED IN AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN 5) ALARM WAS IDENTIFIED IN THE LOG. THE ALARM OCCURRED ON (B)(6) 2014 AT 11:41:19. THIS MEETS INCREASED INTRA PERITONEAL VOLUME (IIPV) CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310863 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1 65 YR