HOMECHOICE
Report
- Report Number
- 1416980-2014-17118
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Date of Event
- April 5, 2014
- Report Date
- May 2, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. ALL RETURNED HOMECHOICE (HC) DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE DEVICE PASSED RITE TESTING. NO ABNORMALITIES WERE IDENTIFIED DURING INTERNAL/EXTERNAL INSPECTIONS. THE PROBLEM OF A HIGH DRAIN ALARM WAS IDENTIFIED IN AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN 5) ALARM WAS IDENTIFIED IN THE LOG. THE ALARM OCCURRED ON (B)(6) 2014 AT 11:41:19. THIS MEETS INCREASED INTRA PERITONEAL VOLUME (IIPV) CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310863 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |