SMALL BATTERY DRIVE
Report
- Report Number
- 8030965-2014-10254
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Date of Event
- May 9, 2014
- Report Date
- May 9, 2014
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO DEVICE BEING WORN DUE TO NORMAL USE AND SERVICING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE EVALUATION WAS CORRECTED. SEE CORRECTED EVALUATION. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD A WORN COUPLING HEAD, VIBRATION AND WAS MAKING NOISE. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WORN BEARINGS FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
IT WAS REPORTED THAT DURING ENGINEERING EVALUATION, IT WAS DISCOVERED THAT THE SMALL BATTERY DRIVE DEVICE WAS MAKING NOISE. THE EVENT WAS NOT RELATED TO SURGERY. THERE WERE NO INJURIES OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 310370 | SMALL BATTERY DRIVE | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT | HWE | SYNTHES OBERDORF | 3201 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |