FDA Adverse Event Malfunction Summary report: N

TENKU DILATATION CATHETER

MDR report key: 3833662 · Received May 27, 2014

Report

Report Number
2024168-2014-03350
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 24, 2014
Report Date
May 2, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
LOX
PMA / PMN Number
K103110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AGE, ESTIMATED: REPORTED AS 60'S. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. A QUERY OF THE COMPLAINT HANDLING DATABASE REVEALED NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. THE TENKU DILATION CATHETER DEVICE IS AN ABBOTT VASCULAR MANUFACTURED DEVICE WHICH IS DISTRIBUTED BY (B)(4). THOUGH THIS DEVICE IS NOT COMMERCIALLY AVAILABLE FOR SALE IN THE US, IT IS SIMILAR TO A DEVICE CURRENTLY MARKETED FOR SALE IN THE US.

Description of Event or Problem · 1

IT WAS REPORTED THAT USING A RADIAL ARTERY ACCESS APPROACH DURING A PROCEDURE OF THE HEAVILY CALCIFIED, MILDLY TORTUOUS, 90% STENOSED, MID LEFT ANTERIOR DESCENDING (LAD) ARTERY THE 2.5 X 15 MM TENKU BALLOON DILATATION CATHETER (BDC) WAS BEING INFLATED ONCE WHEN THE BALLOON RUPTURED AT 6 ATMOSPHERE AFTER 10 SECONDS. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT. THERE WAS NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310023 TENKU DILATATION CATHETER CORONARY DILATATION CATHETER LOX AV-TEMECULA-CT 40301G1

Patients

Seq Age Sex Outcome Treatment
1 60 YR GUIDE WIRE: SION BLUE