FDA Adverse Event Malfunction Summary report: N

LIGAMAX CLIP APPLIER

MDR report key: 3833652 · Received May 27, 2014

Report

Report Number
3005075853-2014-03569
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 18, 2014
Report Date
May 7, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4) = BENT ADVANCER, MALFORMED CLIP, INDICATOR WHEEL FAILURE, CLIP GAP. THE ANALYSIS RESULTS FOUND THAT ONE EL5ML DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE EVENT REPORTED, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS FIRED AND DURING THE FIRST FIRING SEQUENCE 2 CLIPS WERE FED AT THE SAME TIME; 2 PEAR SHAPED CLIPS WERE RELEASED. NEXT FIRING SEQUENCE THE CLIP WAS NOT PROPERLY FED CAUSING 1 GAP CLIP. THE DEVICE WAS NOTED TO HAVE THE TIP OF THE ADVANCER BENT. A POSSIBLE CAUSE FOR THE CONDITION OF BENT ADVANCER IS ACTUATING THE DEVICE WHEN THE JAWS ARE NOT CLEAR OF THE TROCAR. ACTUATING THE DEVICE WITH THE JAWS CLOSED MAY BEND THE ADVANCER. PLEASE NOTE THAT PRIOR TO LOADING A CLIP IN THE JAWS, ENSURE THAT THE DEMARCATION BETWEEN THE JAWS AND THE DEVICE SHAFT IS PAST THE END OF THE TROCAR CANNULA. ADDITIONALLY, EXCESSIVELY APPLYING A SIDE LOAD TO THE JAWS, CAUSING THEM TO PARTIALLY COLLAPSE COULD RESULT IN A CLIP MALFORMATION. THE DEVICE JAWS SHOULD BE FULLY OPEN AND PARALLEL UPON INITIATING THE FIRING OF THE DEVICE. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED BUT THE ORANGE INDICATOR WAS FOUND UNDERTRAVELED. PLEASE NOTE THAT THIS FINDING IS NOT RELATED WITH THE INCIDENT REPORTED. IT SHOULD BE NOTED THAT AS PART OF OUR QUALITY PROCESS, EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIP WAS MALFORMED. THE DEVICE WAS USED FOR THE CYSTIC ARTERY. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.