FDA Adverse Event Malfunction Summary report: N

PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL)

MDR report key: 3833625 · Received May 27, 2014

Report

Report Number
1031452-2014-01757
Event Type
Malfunction
Date Received
May 27, 2014
Report Date
April 15, 2014
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

DEALER ALLEGES THE ON AND OFF SWITCH NOT ALARMING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310448 PORTABLE OXYGEN GENERATOR (CONCENTRATOR, HOMEFILL) 868.5440 CAW INVACARE FLORIDA OPERATIONS IRC10LXO2

Patients

Seq Age Sex Outcome Treatment
1 Other