FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 3833491 · Received May 27, 2014

Report

Report Number
3004209178-2014-85038
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 6, 2014
Report Date
May 6, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

EVALUATED ONE OPENED AND USED RESERVOIR. CHECKED O-RINGS AND STOPPER GROOVE FOR DEFECTS AND NONE WERE FOUND. RAN BASAL AND BOLUS LEAKING TEST. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIR PASSED PER INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT INSULIN WAS LEAKING INTO THE RESERVOIR CHAMBER. THE CUSTOMER'S BLOOD GLUCOSE READING WAS 385 MG/DL. THE CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD BEEN EXPOSED TO FLUID. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310207 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A H8161543

Patients

Seq Age Sex Outcome Treatment
1 19 YR