FDA Adverse Event Malfunction Summary report: N

PROGRASP FORCEPS INSTRUMENT

MDR report key: 3833200 · Received May 27, 2014

Report

Report Number
2955842-2014-03243
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 2, 2014
Report Date
May 12, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS FRAYED. THE PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY WEAR. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DA VINCI PROSTATECTOMY SURGICAL PROCEDURE, IT WAS NOTED THAT BROKEN WIRES WERE OBSERVED ON THE PROGRASP FORCEPS INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311226 PROGRASP FORCEPS INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-11 M10140130 906

Patients

Seq Age Sex Outcome Treatment
1