PROGRASP FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-03243
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Date of Event
- May 2, 2014
- Report Date
- May 12, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS FRAYED. THE PITCH CABLE WAS FRAYED AT THE PROXIMAL CLEVIS HUB. THE CLEVIS DID NOT EXHIBIT ANY WEAR. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT AFTER A DA VINCI PROSTATECTOMY SURGICAL PROCEDURE, IT WAS NOTED THAT BROKEN WIRES WERE OBSERVED ON THE PROGRASP FORCEPS INSTRUMENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.