FDA Adverse Event Injury Summary report: N

OPTIBOND XTR

MDR report key: 3833185 · Received May 27, 2014

Report

Report Number
2024312-2014-00393
Event Type
Injury
Date Received
May 27, 2014
Report Date
April 16, 2014
Manufacturer
KERR CORPORATION
Product Code
KLE
PMA / PMN Number
K101423
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

IT WAS DISCOVERED ON (B)(6) 2014 THAT THE DOCTOR HAD NOT REFRIGERATED THE OPTIBOND XTR PRODUCT, WHICH IS REQUIRED AS PER THE DIRECTIONS FOR USE. IT WAS INITIALLY REPORTED THAT THE PRODUCT INVOLVED IN THE ALLEGED INCIDENT WAS NOT RETURNED AND EVALUATIONS WERE PERFORMED ON A RETAINED SAMPLE; HOWEVER, THE OPTIBOND XTR PRODUCT (ADHESIVE (CATALOG #35108, LOT #4674868 AND PRIMER (CATALOG # 35107 AND LOT #4657224)) INVOLVED IN THE ALLEGED INCIDENT HAD BEEN RETURNED AND THE VISUAL EVALUATION WAS PERFORMED ON THE RETURNED SAMPLE, WHICH YIELDED RESULTS WITHIN SPECIFICATIONS. A DHR REVIEW INDICATED THAT THERE WERE NO DEVIATIONS FROM THE MANUFACTURING PROCESS. IN ADDITION, NO SIMILAR COMPLAINTS WERE RECEIVED WITH REGARD TO THESE LOTS.

Additional Manufacturer Narrative · 1

SPECIFIC PATIENT INFORMATION WITH REGARD TO GENDER, AGE, AND WEIGHT WAS NOT PROVIDED. THE DOCTOR REPORTED THAT THE PATIENT HAD CONTACTED THE OFFICE WITHIN APPROXIMATELY THREE (3) DAYS OF PLACEMENT REPORTING SENSITIVITY ISSUES. THE DOCTOR HAD THE PATIENT RETURN TO THE OFFICE AND TOOK AN X-RAY; HOWEVER, NO ISSUES WITH THE RESTORATION WERE NOTED. THE DOCTOR REQUESTED THAT THE PATIENT WAIT 2-3 WEEKS TO SEE IF THE SENSITIVITY WOULD SUBSIDE; HOWEVER, AFTER WAITING THE REQUESTED TIME, THE SENSITIVITY WAS STILL PRESENT. THE DOCTOR HAD THE PATIENT RETURN TO THE OFFICE AND REPLACED THE PATIENT'S RESTORATIONS. NO TREATMENT DATES WERE PROVIDED. THE DOCTOR REPORTED THAT THE SENSITIVITY HAS BEEN REDUCED; HOWEVER, THE PATIENT IS STILL EXPERIENCING SOME SENSITIVITY. THE OPTIBOND XTR PRODUCT (ADHESIVE (CATALOG #35108, LOT #4674868) AND PRIMER (CATALOG # 35107 AND LOT #4657224)) INVOLVED IN THE ALLEGED INCIDENT WAS NOT RETURNED; THEREFORE, A VISUAL EVALUATION WAS PERFORMED ON A RETAINED SAMPLE, YIELDING RESULTS WITHIN SPECIFICATIONS. A DHR REVIEW INDICATED THAT THERE WERE NO DEVIATIONS FROM THE MANUFACTURING PROCESS. IN ADDITION, NO SIMILAR COMPLAINTS WERE RECEIVED WITH REGARD TO THIS LOT.

Description of Event or Problem · 1

A DOCTOR ALLEGED THAT ONE HUNDRED AND TWENTY (120) PATIENTS HAD EXPERIENCED SENSITIVITY AFTER RESTORATIONS WERE PLACED USING THE OPTIBOND XTR PRODUCT. THIS IS THE NINETY-FIFTH OF ONE HUNDRED AND TWENTY (120) REPORTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311221 OPTIBOND XTR RESIN TOOTH BONDING AGENT KLE KERR CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 Other| R (B)(4)| SONICFILL