FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3833182 · Received May 27, 2014

Report

Report Number
1416980-2014-17024
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 4, 2014
Report Date
May 4, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. WHILE THE DEVICE ANALYSIS COULD NOT BE COMPLETED, A LOOSE CONNECTION AND DISCONNECTION WAS REPORTED WHICH IS A KNOWN CAUSE OF THIS ALARM. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET) ALARM OCCURRED DURING DRAIN ONE ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THE CARE GIVER (CG) INDICATED THAT THE PATIENT LINE EXTENSION BECAME SEPARATED FROM THE CONNECTION POINT WITH THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO END THERAPY AND START OVER WITH NEW SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311689 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 65 YR HOMECHOICE