PLUM A+3 V11.51 REFR
Report
- Report Number
- 9615050-2014-02245
- Event Type
- Malfunction
- Date Received
- March 25, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 24, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K052052
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS REC'D. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT CHANNELS 1, 2, AND 3 OF THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL ENGINEERING DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "E301 DURING INFUSION". NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, CHANNELS 1, 2, AND 3 OF THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173697 | PLUM A+3 V11.51 REFR | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |