SYMBIQ SINGLE CHANNEL

MDR report key: 3833070 · Received March 25, 2014

Report

Report Number
9615050-2014-02269
Event Type
Malfunction
Date Received
March 25, 2014
Date of Event
January 1, 2014
Report Date
February 24, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K1109010
Removal / Correction Number
Z-0069-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THAT THE DEVICE PASSED TESTING; HOWEVER, A REVIEW OF THE DEVICE HISTORY INDICATED TOUCHSCREEN FAIL ERRORS. ADDITIONAL TESTING FOUND FLUID INGRESS ON THE BOTTOM OF THE TOUCHSCREEN WHICH WAS THE PROBABLE CAUSE OF THE CUSTOMER'S REPORTED NON RESPONSIVE TOUCHSCREEN. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT THE DEVICE TOUCHSCREEN DOES NOT RESPOND WHEN PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE TOUCHSCREEN WAS UNRESPONSIVE AT START UP. NO ADDITIONAL INFO WAS PROVIDED.