SYMBIQ DUAL CHANNEL
Report
- Report Number
- 9615050-2014-02268
- Event Type
- Malfunction
- Date Received
- March 25, 2014
- Date of Event
- January 1, 2014
- Report Date
- February 24, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY, DURING TESTING, THE DEVICE ALARMED WITH A S233 (OVER TEMPERATURE-PSC) ALARM CODE. THE PROBABLE CAUSE OF THE S233 MALFUNCTION ALARM CODE WAS A BROKEN FAN. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THAT THE DEVICE ALARMED WITH MULTIPLE S233 (OVER TEMPERATURE - PSC) MALFUNCTION ALARM CODES. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR AN UNSPECIFIED REASON. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS AND NO REPORTED DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH A S233 MALFUNCTION ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED.