PLUM XLM W/DATAPORT
Report
- Report Number
- 9615050-2014-02262
- Event Type
- Malfunction
- Date Received
- March 25, 2014
- Date of Event
- January 1, 2014
- Report Date
- March 3, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K010924
- Removal / Correction Number
- FA311-04
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS MISSING. THE CUSTOMER CONTACT'S REPORT OF DEVICE WILL NOT TURN ON IN BATTERY MODE WAS DUPLICATED DURING TESTING. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS MISSING. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A REPORT THAT THE DEVICE WILL NOT TURN ON IN BATTERY MODE. THIS IS NOT A REPORTABLE MALFUNCTION. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 173735 | PLUM XLM W/DATAPORT | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |