FRESENIUS 2008K2 MACHINE

MDR report key: 3833023 · Received March 25, 2014

Report

Report Number
2937457-2014-00449
Event Type
Malfunction
Date Received
March 25, 2014
Date of Event
February 24, 2014
Report Date
March 1, 2014
Manufacturer
CONCORD MANUFACTURING
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER FACILITY REPORTED A SALINE BAG BACK FILED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THE MACHINE WAS SERVICE AND THE AIR SEPARATOR REPLACED DUE TO A CRACK.