GEMSTAR PAIN MGE SNG
MDR report key: 3832956
·
Received March 26, 2014
Report
- Report Number
- 9615050-2014-02332
- Event Type
- Malfunction
- Date Received
- March 26, 2014
- Date of Event
- February 21, 2014
- Report Date
- February 23, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1159-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING TESTING, AT POWER UP THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED TO THE SERVICE CENTER BY THE CUSTOMER CONTACT FOR UPDATE/RECERTIFICATION. NO ADDITIONAL INFORMATION WAS PROVIDED. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER UPON POWER ON, THE DEVICE ALARMED WITH A 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE.