FDA Adverse Event
Malfunction
Summary report: N
7900
MDR report key: 3832945
·
Received March 17, 2014
Report
- Report Number
- 9680959-2014-00362
- Event Type
- Malfunction
- Date Received
- March 17, 2014
- Date of Event
- March 3, 2014
- Report Date
- March 17, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS GMBH
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE BOARDS WERE RESEATED AND THE POWER PLUG WAS REPAIRED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP AND SHUT DOWN. THIS CAUSED AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 157270 | 7900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS GMBH | 7900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |