FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3832844 · Received March 17, 2014

Report

Report Number
9615050-2014-02024
Event Type
Malfunction
Date Received
March 17, 2014
Date of Event
February 19, 2014
Report Date
February 19, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1159-2014
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND THE DEVICE ALARMED FOR SERVICE ALARM CODE 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY). THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 COLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM THE PRE-OPERATIVE DEPARTMENT WITH A NOTE THAT STATED, "ALARM CAME UP". THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156827 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK