FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3832755 · Received May 27, 2014

Report

Report Number
2955842-2014-03225
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 8, 2014
Report Date
May 8, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS INVESTIGATION FOUND THE INSTRUMENT'S PITCH CABLE WAS FRAYED. THE CONDUCTOR WIRES WERE INTACT AND UNDAMAGED, BUT THE DRIVE CABLE WAS FRAYED. THE PITCH UP CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, IT WAS NOTED THAT A FRAYED WIRE WAS OBSERVED ON THE MARYLAND BIPOLAR FORCEPS INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO THE PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310490 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 N10131123 038

Patients

Seq Age Sex Outcome Treatment
1