FDA Adverse Event Malfunction Summary report: N

PLM A+ SPANISH 110V

MDR report key: 3832712 · Received March 26, 2014

Report

Report Number
9615050-2014-02290
Event Type
Malfunction
Date Received
March 26, 2014
Date of Event
January 1, 2014
Report Date
February 27, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER PIN WAS BROKEN. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH A NOTE THAT STATED, "DOOR BROKEN." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176696 PLM A+ SPANISH 110V 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK PLUM A+ SOFTWARE MODULE, LIST #12097,| SN (B)(4)