GEMSTAR PAIN MGE SNG
Report
- Report Number
- 9615050-2014-02278
- Event Type
- Malfunction
- Date Received
- March 26, 2014
- Date of Event
- January 23, 2014
- Report Date
- February 6, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1169-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
DURING TESTING AT THE SERVICE CENTER, THE DEVICE WOULD NOT POWER ON DC POWER; HOWEVER, WAS ABLE TO BE POWERED ON USING AC POWER. DURING FURTHER TESTING CORROSION DUE TO LEAKAGE FROM THE DISPOSABLE BATTERIES WAS NOTED ON THE BATTERY CONTACT, BATTERY DOOR, BATTERY COMPARTMENT, AND PRINTED CIRCUIT BOARD. THIS DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT THAT THE DEVICE WOULD NOT POWER ON USING DC POWER; HOWEVER, IT WOULD POWER ON USING THE POWER CORD. THIS IS NOT A REPORTABLE FUNCTION. DURING VERIFICATION TESTING AT THE SERVICE CENTER, LEAKAGE FROM THE DISPOSABLE BATTERIES WAS NOTED IN THE BATTERY COMPARTMENT OF THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176345 | GEMSTAR PAIN MGE SNG | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |