FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3832686
·
Received March 27, 2014
Report
- Report Number
- 1720753-2014-02702
- Event Type
- Malfunction
- Date Received
- March 27, 2014
- Date of Event
- March 5, 2014
- Report Date
- March 27, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE TANK AND THE SNUBBER BOARD WERE REPLACED AND THE FILAMENT WAS CALIBRATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A HIGH MA ERROR MESSAGE DURING A PROCEDURE WITH PATIENT INVOLVEMENT, THEREFORE THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ARO (ACCIDENTAL RADIATION OCCURRENCE). THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179459 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |