FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3832664
·
Received March 27, 2014
Report
- Report Number
- 1720753-2014-02716
- Event Type
- Malfunction
- Date Received
- March 27, 2014
- Date of Event
- March 6, 2014
- Report Date
- March 27, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE RTOS WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179388 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |