FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3832641 · Received March 27, 2014

Report

Report Number
1720753-2014-02719
Event Type
Malfunction
Date Received
March 27, 2014
Date of Event
March 7, 2014
Report Date
March 27, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE PRIMARY SIDE OF TRANSFORMER WAS EVALUATED AND RESTRAPPED. THE SYSTEM SOFTWARE AND CALIBRATION FILES WERE RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM LOCKED UP DURING USE. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179341 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1