FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3832640
·
Received March 27, 2014
Report
- Report Number
- 1720753-2014-02721
- Event Type
- Malfunction
- Date Received
- March 27, 2014
- Date of Event
- March 4, 2014
- Report Date
- March 27, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE RIBBON CABLE CONNECTING P3 ON VIDEO CONTROLLER PCB WAS EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THEY WERE UNABLE TO VIEW FLUOROSCOPIC IMAGE BECAUSE THE SYSTEM DISPLAYED A BLACK IMAGE. THIS ISSUE WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO THE INABILITY TO SEE THE LIVE IMAGE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179595 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |