FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 3832578
·
Received March 27, 2014
Report
- Report Number
- 1828100-2014-00203
- Event Type
- Malfunction
- Date Received
- March 27, 2014
- Date of Event
- March 5, 2014
- Report Date
- March 6, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEM CORP
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
WHEN THE FSR WENT OUT FOR REPAIR HE FOUND THAT THE HOSE WAS COLLAPSING IN THIGH SPEED.
Description of Event or Problem · 1
IT WAS REPORTED THAT CARDIOPLEGIA FLOW WAS NOT WORKING IN SPEED 3 FROM LARGE TO SMALL TANK ON THE COOLER HEATER UNIT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179508 | TERUMO TCM II COOLING AND HEATING SYSTEM | TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEM CORP | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |