FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3832458 · Received May 27, 2014

Report

Report Number
1416980-2014-17002
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. THIS COMPLAINT IS FOR A REPORT OF A SYSTEM ERROR (SE) 2240 DURING THE INITIAL DRAIN, WHERE THE CARE GIVER (CG) STATED THE PATIENT LINE WAS NOT FULLY PRIMED BEFORE THE PATIENT CONNECTED TO THE SET UP. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. SECTION ¿PREPARE FOR THERAPY¿ PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. SECTION ¿PREPARE FOR THERAPY¿ PROVIDES STEP-BY-STEP INSTRUCTIONS FOR WHEN AND HOW TO CONNECT TO THE DISPOSABLE SET. SECTION ¿TROUBLESHOOTING¿ PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY RE-PRIMING THE PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 (AIR IN SET) ALARM OCCURRED DURING THE INITIAL DRAIN ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE EVENT. THE CARE GIVER (CG) STATED THE PATIENT LINE WAS NOT PRIMED FULLY BEFORE THE HP CONNECTED TO THE SET UP. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP TO START OVER WITH NEWS SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311300 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 68 YR HOMECHOICE