FDA Adverse Event
Malfunction
Summary report: N
2008K HEMODIALYSIS SYS OLC/DIASAFE PLUS
MDR report key: 3832431
·
Received March 27, 2014
Report
- Report Number
- 2937457-2014-00473
- Event Type
- Malfunction
- Date Received
- March 27, 2014
- Date of Event
- March 6, 2014
- Report Date
- March 6, 2014
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- KDI
- PMA / PMN Number
- K994267
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MFR VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.
Description of Event or Problem · 1
A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE NO PARTS REPLACED ON THE DEVICE BY THE TECHNICIAN AND THE MACHINE WAS RETURNED TO SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 179485 | 2008K HEMODIALYSIS SYS OLC/DIASAFE PLUS | KDI | FRESENIUS MEDICAL CARE NORTH AMERICA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK SALINE (DISCARDED-NOT USED) |