FDA Adverse Event Malfunction Summary report: N

2008K HEMODIALYSIS SYS OLC/DIASAFE PLUS

MDR report key: 3832418 · Received March 27, 2014

Report

Report Number
2937457-2014-00479
Event Type
Malfunction
Date Received
March 27, 2014
Date of Event
November 27, 2013
Report Date
March 19, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MFR VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE NO PARTS REPLACED ON THE DEVICE BY THE TECHNICIAN AND THE MACHINE WAS RETURNED TO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
179549 2008K HEMODIALYSIS SYS OLC/DIASAFE PLUS KDI FRESENIUS MEDICAL CARE NORTH AMERICA

Patients

Seq Age Sex Outcome Treatment
1 UNK SALINE (DISCARDED-NOT USED)