FDA Adverse Event Malfunction Summary report: N

VASOVIEW HEMOPRO 2

MDR report key: 3832391 · Received February 10, 2014

Report

Report Number
2242352-2014-00076
Event Type
Malfunction
Date Received
February 10, 2014
Date of Event
January 15, 2014
Report Date
January 15, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K101274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED SIGNS OF CLINICAL USAGE AND EVIDENCE OF BLOOD. A VISUAL INSPECTION DETERMINED THAT THE HOT JAW WAS BENT AND THE WHITE WIRE WELDER TO THE JAW WAS EXPOSED. ADDITIONAL EVALUATION IS REQUIRED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE JAWS CRACKED ON THE VASOVIEW HEMOPRO 2. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88111 VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-4000 25088644

Patients

Seq Age Sex Outcome Treatment
1 64 YR