FDA Adverse Event Malfunction Summary report: N

VASO VIEW HEMOPRO

MDR report key: 3832383 · Received February 10, 2014

Report

Report Number
2242352-2014-00086
Event Type
Malfunction
Date Received
February 10, 2014
Report Date
January 16, 2014
Manufacturer
MAQUET CARDIOVASCULAR, LLC
Product Code
GEI
PMA / PMN Number
K052274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. A VISUAL INSPECTION DID NOT IDENTIFY ANY NON-CONFORMITIES. A FUNCTIONAL TEST WAS CONDUCTED ON THE DISSECTION TIP. THE DEVICE WAS ATTACHED TO A REFERENCE ENDOSCOPE AND VIEWED ON A MONITOR; VISIBILITY WAS NOT CLEAR THROUGH THE DISSECTION TIP, DISTORTION WAS OBSERVED. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. ADDITIONAL INFORMATION IS REQUIRED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).

Description of Event or Problem · 1

THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE DISSECTION TIP ON THE VASO VIEW HEMOPRO WAS CLOUDY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87601 VASO VIEW HEMOPRO ENDOSCOPIC VESSEL HARVESTING GEI MAQUET CARDIOVASCULAR, LLC VH-3000 25089465

Patients

Seq Age Sex Outcome Treatment
1 NI