VASO VIEW HEMOPRO
Report
- Report Number
- 2242352-2014-00086
- Event Type
- Malfunction
- Date Received
- February 10, 2014
- Report Date
- January 16, 2014
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE DEVICE WAS RETURNED TO THE FACTORY FOR EVALUATION. IT SHOWED NO SIGNS OF CLINICAL USAGE OR EVIDENCE OF BLOOD. A VISUAL INSPECTION DID NOT IDENTIFY ANY NON-CONFORMITIES. A FUNCTIONAL TEST WAS CONDUCTED ON THE DISSECTION TIP. THE DEVICE WAS ATTACHED TO A REFERENCE ENDOSCOPE AND VIEWED ON A MONITOR; VISIBILITY WAS NOT CLEAR THROUGH THE DISSECTION TIP, DISTORTION WAS OBSERVED. BASED UPON THE RECEIVED CONDITION OF THE DEVICE, THE REPORTED COMPLAINT WAS CONFIRMED. ADDITIONAL INFORMATION IS REQUIRED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WAS NO NON-CONFORMANCE RECORDED IN THE LOT HISTORY. (B)(4).
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE DISSECTION TIP ON THE VASO VIEW HEMOPRO WAS CLOUDY. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87601 | VASO VIEW HEMOPRO | ENDOSCOPIC VESSEL HARVESTING | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25089465 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |