ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-14812
- Event Type
- Malfunction
- Date Received
- May 27, 2014
- Report Date
- May 21, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: PUMP WAS EXERCISED FOR 24 HOURS; BATTERY WAS REMOVED FOR 6 HOURS AND THE TIME AND DATE RESET TO DEFAULT. THERE'S A CRACK ALONG THE BATTERY COMPARTMENT THREADS. PUMP WAS OPENED TO INVESTIGATE AND THE INTERNAL BATTERY WAS FOUND TO BE LEAKING AND UNABLE TO HOLD A CHARGE. PUMP BOOTED TO THE VERIFY SCREEN WITH DIM PINK CONTRAST. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT, LEAKING INTERNAL BATTERY, AND DIM PINK DISPLAY. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2014.