PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2014-01055
- Event Type
- Malfunction
- Date Received
- February 10, 2014
- Date of Event
- January 17, 2014
- Report Date
- January 17, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. ADDITIONALLY DURING TESTING, THE DEVICE ALARMED WITH AN N250 (DOOR OPEN WHILE PUMPING) ALARM CONDITION. THIS WAS DUE TO A DEFECTIVE FLUID SHIELD. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN, THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPT FOR A REPORT OF BROKEN DOOR ROLLER AND N250. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87603 | PLUM A+ DRIVER ED 2 | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE, LIST#20791, SN UNK |