FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3832273 · Received May 27, 2014

Report

Report Number
1416980-2014-16990
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
February 3, 2014
Report Date
May 5, 2014
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A SYSTEM ERROR 2240 ALARM (AIR IN SET) WAS IDENTIFIED IN THE LOG WHICH INDICATES A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. THE ALARM OCCURRED ON (B)(6) 2014 AT 06:27:18. NO ADDITIONAL INFORMATION IS AVAILABLE.