FDA Adverse Event Summary report: N

VERSAFITCUP CC CANCELLOUS BONE SCREW

MDR report key: 3832263 · Received May 14, 2014

Report

Report Number
3005180920-2014-00055
Date Received
May 14, 2014
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
LZO
PMA / PMN Number
K103352
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

DOCUMENT REVIEW: VERSAFITCUP CC CANCELLOUS BONE SCREW L 35 0 6.5, CODE 01.26.65.35 - LOT 121042 ((B)(4) SCREWS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE (B)(4) ITEMS BELONGING TO THIS LOT HAVE BEEN ALREADY SOLD AND NO SIMILAR INCIDENTS HAVE BEEN REPORTED. VERSAFITCUP CC CANCELLOUS BONE SCREW L 30 0 6.5, CODE 01.26.65.30 - LOT 131304 ((B)(4) SCREWS PRODUCED): ALL PARAMETERS WERE FOUND TO BE IN ACCORDANCE WITH THE SPECIFICATIONS VALID AT THE TIME OF MANUFACTURING. THE (B)(4) ITEMS BELONGING TO THIS LOT HAVE BEEN ALREADY SOLD AND NO SIMILAR INCIDENTS HAVE BEEN REPORTED. IT WAS REPORTED THAT THE SURGEON DID NOT PRE-DRILL A HOLE IN THE PT BONE, AS CLEARLY REQUESTED IN THE VERSAFITCUP CC TRIO SURGICAL TECHNIQUE (K103352). ON THE BASIS OF THE DATA COLLECTED, THE EVENT IS HIGHLY LIKELY DUE TO THE SURGICAL MISTAKE ABOVE DESCRIBED.

Description of Event or Problem · 1

REF IMP # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
287919 VERSAFITCUP CC CANCELLOUS BONE SCREW ACETABULAR BONE SCREW L 35 0 6.5 LZO MEDACTA INTERNATIONAL, SA

Patients

Seq Age Sex Outcome Treatment
1