FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3832230 · Received March 7, 2014

Report

Report Number
9615050-2014-01788
Event Type
Malfunction
Date Received
March 7, 2014
Date of Event
January 1, 2014
Report Date
February 4, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1159-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, AT POWER UP THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THE CUSTOMER REPORTED ERROR 09/007/036 SERVICE ALARM CODE WAS NOT DUPLICATE; HOWEVER, WAS NOTED IN THE HISTORY. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER WITH A REPORT FROM THE CUSTOMER CONTACT "ERROR 09/007/036". NO ADD'L INFO WAS PROVIDED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER THE DEVICE ALARMED WITH A 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
138505 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA