FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3832229 · Received March 7, 2014

Report

Report Number
9615050-2014-01789
Event Type
Malfunction
Date Received
March 7, 2014
Date of Event
December 18, 2013
Report Date
February 6, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
Z-1159-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TESTING AND INVESTIGATION FOUND AT POWER UP THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM AN UNK FLOOR WITH AN UNSIGNED NOTE INDICATING THE ERROR CODE. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH AN 11/004 SERVICE ALARM CODE. NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
138778 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK