GEMSTR 7 THPY ALT ST
Report
- Report Number
- 9615050-2014-01789
- Event Type
- Malfunction
- Date Received
- March 7, 2014
- Date of Event
- December 18, 2013
- Report Date
- February 6, 2014
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- Z-1159-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND AT POWER UP THE DEVICE ALARMED FOR 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE CODE ALARM. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE ALARMED WITH AN 11/004 (LESS THAN 2.0 VOLTS ON LITHIUM BATTERY) SERVICE ALARM CODE. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FROM AN UNK FLOOR WITH AN UNSIGNED NOTE INDICATING THE ERROR CODE. NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, THE DEVICE ALARMED WITH AN 11/004 SERVICE ALARM CODE. NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 138778 | GEMSTR 7 THPY ALT ST | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |