FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3832192 · Received May 27, 2014

Report

Report Number
2937094-2014-00409
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
March 24, 2014
Report Date
March 24, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP EXHIBITED A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/GLASS CAP FUSION ZONE AT THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE ROTATED INDEPENDENTLY OF METAL CAP; THE METAL CAP EXHIBITED MILD BURNT ON DETRITUS; THE GLASS CAP EXHIBITED SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE OUTER FLOW TUBING EXHIBITED MILD SIGNS OF CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY ALSO EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH LIMITING THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE "FIBER BURNT OUT" AT 41 MINUTES AND 47 SECONDS AND 306,810 JOULES OF USE. THE PROCEDURE WAS COMPLETED USING A SECOND FIBER. PATIENT OUTCOME: "NO INJURY TO THE PATIENT".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310495 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 346A

Patients

Seq Age Sex Outcome Treatment
1