FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION WITH TELEPACK

MDR report key: 3832038 · Received March 7, 2014

Report

Report Number
2221819-2014-00061
Event Type
Malfunction
Date Received
March 7, 2014
Date of Event
February 14, 2014
Report Date
February 14, 2014
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENTS OF THE SYSTEM'S HARD DRIVES, MOTHER BOARD, AND COOLING FAN. SYSTEM WAS CALIBRATED AND SAFETY TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE PANORAMA CENTRAL STATION, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
138540 PANORAMA CENTRAL STATION WITH TELEPACK CENTRAL STATION MONITOR MHX MINDRAY DS USA, INC.