FDA Adverse Event Malfunction Summary report: N

SR90D M-A SCREW ADJUSTMENT TOOL

MDR report key: 3832026 · Received May 27, 2014

Report

Report Number
0009617544-2014-00236
Event Type
Malfunction
Date Received
May 27, 2014
Date of Event
April 30, 2014
Report Date
April 30, 2014
Manufacturer
STRYKER SPINE-FRANCE
Product Code
LXH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: FUNCTIONAL EVALUATION;RESULTS: THE PRODUCT WAS RELEASED FROM MANUFACTURING OVER 11 YEARS AGO. THE INSTRUMENTS HAVE A FINITE LIFE AND ARE SUBJECT TO FATIGUE AND WEAR ASSOCIATED WITH NORMAL USE. THE MOST LIKELY CAUSE OF THE REPORTED FRACTURE IS FATIGUE OF THE INSTRUMENT OVER AN EXTENDED PRODUCT LIFETIME.CONCLUSION: THE SR90D M-A ADJUSTMENT TOOL WAS CONFIRMED TO HAVE A FRACTURE TIP, 1 OF THE 4 PRONGS FRACTURED. THE ISSUE OCCURRED WHILE THE SURGEON WAS REMOVING A SCREW DURING A REVISION SURGERY TO EXTEND THE CONSTRUCT. THE PRODUCT ISSUE CAUSED A 3 MINUTE SURGICAL DELAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DURING 90D REMOVE SURGERY, THE SURGEON USED THE MULTIANGLE ALIGNMENT TOOL FOR REMOVING THE SCREW. HOWEVER THE TIP OF THE MULTIANGLE ALIGNMENT TOOL BROKE. THEREFORE THE SURGEON TRIED TO REMOVE SCREW USING THE MULTIANGLE SCREW INSERTER. HOWEVER THE TIP OF THE MULTIANGLE SCREW INSERTER WAS DEFORMING. THE SURGEON USED THE BROKEN MULTIANGLE ALIGNMENT TOOL, AND REMOVED SCREW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DURING 90D REMOVE SURGERY, THE SURGEON USED THE MULTIANGLE ALIGNMENT TOOL FOR REMOVING THE SCREW. HOWEVER THE TIP OF THE MULTIANGLE ALIGNMENT TOOL BROKE. THEREFORE THE SURGEON TRIED TO REMOVE SCREW USING THE MULTIANGLE SCREW INSERTER. HOWEVER THE TIP OF THE MULTIANGLE SCREW INSERTER WAS DEFORMING. THE SURGEON USED THE BROKEN MULTIANGLE ALIGNMENT TOOL, AND REMOVED SCREW.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
310803 SR90D M-A SCREW ADJUSTMENT TOOL INSTRUMENT LXH STRYKER SPINE-FRANCE 022731

Patients

Seq Age Sex Outcome Treatment
1