FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3831972 · Received May 27, 2014

Report

Report Number
2531779-2014-14799
Event Type
Malfunction
Date Received
May 27, 2014
Report Date
May 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. THE INVESTIGATION FOUND A CRACK IN THE BATTERY COMPARTMENT EXTENDING FROM THE CASE SEAL TO THE GRIP PAD. ON INVESTIGATION, THE PUMP POWERED UP AND ALL THE KEYPAD BUTTONS FUNCTIONED PROPERLY. REMOVAL OF THE KEYPAD COVER FOUND CONTAMINATION UNDER ALL THE BUTTON CONTACTS. UNRELATED TO THE CASING AND CONTAMINATION ISSUES, IT WAS OBSERVED THAT THE KEYPAD COVER WAS TORN. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION FOUND THAT THE BATTERY COMPARTMENT WAS CRACKED AND THERE WAS CONTAMINATION UNDER THE KEYPAD BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
311731 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 69 YR