FDA Adverse Event
Malfunction
Summary report: N
TERUMO TCM II COOLING AND HEATING SYSTEM
MDR report key: 3831971
·
Received March 7, 2014
Report
- Report Number
- 1828100-2014-00136
- Event Type
- Malfunction
- Date Received
- March 7, 2014
- Date of Event
- February 12, 2014
- Report Date
- February 13, 2014
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DWC
- PMA / PMN Number
- K883603
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
TECHNICAL SUPPORT SUGGESTED TO REPLACE THE CARDIOPLEGIA SENSOR. THE BIOMED IS REPAIRING THE UNIT AT THE HOSPITAL AND IS NOT REQUESTING SERVICE AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE CARDIOPLEGIA DID NOT COOL LOW ENOUGH. THE USER FACILITY'S BIOMEDICAL ENGINEER (BIOMED) REPORTED WHEN HE CHECKED THE UNIT ON (B)(6) 2014, THE CARDIOPLEGIA DISPLAY READ 28 DEGREES CELSIUS AND THE OUTPUT LINE IN THE SMALL TANK READ 14.1 DEGREES CELSIUS. THE DEVICE WAS NOT CHANGED OUT. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO DELAYS, NO BLOOD LOSS, OR NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 139066 | TERUMO TCM II COOLING AND HEATING SYSTEM | TCM II COOLING AND HEATING SYSTEM | DWC | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 4415 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |