FDA Adverse Event Malfunction Summary report: N

DAYTONA TABLE; TAVO-VL4

MDR report key: 3831961 · Received March 7, 2014

Report

Report Number
3008478792-2014-00001
Event Type
Malfunction
Date Received
March 7, 2014
Date of Event
February 7, 2014
Report Date
March 7, 2014
Manufacturer
CONSULTRONIX S.A.
Product Code
HRJ
PMA / PMN Number
EXEMPT
Removal / Correction Number
3008478792-3/7/2014-001-
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

A DISTRIBUTOR REPORTED THAT THE DEVICE CONTINUED TO DESCEND PAST THE ZERO POSITION WHEN THE BUTTON WAS PUSHED. THE TABLE DID NOT MOVE SPONTANEOUSLY, BUT COULD BE FORCED BELOW THE ZERO POINT WHEN THE BUTTON WAS PUSHED. UNDER ORDINARY OPERATION, ONCE THE TABLE REACHED THE ZERO POSITION ON THE COLUMN, THE TABLE SHOULD NOT RESPOND TO PUSHING THE BUTTON TO DESCEND. THE WORST CASE SCENARIO FOR RISK IS WHERE THE USER CONTINUES TO FORCE THE TABLE TO DESCEND, WHICH COULD EVENTUALLY SEVER THE ELECTRICAL CABLE / DAMAGE THE ELECTRICAL PLUG. IF THE CABLE IS SEVERED, THERE IS A RISK OF A SHOCK BEING DELIVERED TO THE PATIENT OR USER FROM THE EXPOSED CABLE. THE COMPANY'S INVESTIGATION CONCLUDED THAT THIS CONDITION COULD OCCUR UNDER THE FOLLOWING CIRCUMSTANCES WHICH WOULD MAKE THE POSSIBILITY OF SERIOUS INJURY EXTREMELY REMOTE: A GREATER THAN 15% DROP IN VOLTAGE DELIVERED TO THE UNIT; THE USER PUSHES THE BUTTON TO DRIVE THE UNIT BELOW THE ZERO POINT; THE USER DOES NOT NOTICE THAT THE TABLE CONTINUES TO DESCEND PAST THE ZERO POINT; AND THE USER DELIVERS ENOUGH FORCE (IN A SINGLE INSTANCE OR OVER TIME) TO SEVER THE CABLE OR DAMAGE THE PLUG (APPROXIMATELY 3400N) (NOTE THAT THE COLUMN'S MAXIMUM DOWNWARD FORCE IS 2500N). WHILE THE COMPANY HAS NOT HAD A REPORT OF SHOCK ASSOCIATED WITH THIS FAILURE MODE AND BELIEVES THE POTENTIAL FOR INJURY IS HIGHLY UNLIKELY, THE COMPANY IS SUBMITTING THIS REPORT OUT OF AN ABUNDANCE OF CAUTION. CONSULTRONIX IS ISSUING A FIELD NOTIFICATION TO INFORM CUSTOMERS OF THIS ISSUE AND AN ALUMINUM BLOCK THAT WILL PROVIDE A MECHANICAL STOP TO PROHIBIT THE DESCENT OF THE TABLE PAST THE ZERO POINT ON THE COLUMN. THE FIELD NOTIFICATION AND ALUMINUM BLOCK WILL BE SENT TO ALL CUSTOMERS IN THE U.S. WITH POTENTIALLY AFFECTED PRODUCT.

Description of Event or Problem · 1

A DISTRIBUTOR REPORTED THAT WHEN THE BUTTON WAS PUSHED, THE TABLE WAS ABLE TO MOVE BELOW THE ZERO POINT WHEN DESCENDING ON THE COLUMN. THERE WAS NO REPORT OF A PATIENT BEING INVOLVED IN EVENT, NOR WAS THERE ANY REPORT OF A PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
138869 DAYTONA TABLE; TAVO-VL4 TABLE, INSTRUMENT, POWERED, OPHTHALM HRJ CONSULTRONIX S.A. TAVO-VL4

Patients

Seq Age Sex Outcome Treatment
1