INRATIO2
Report
- Report Number
- 2027969-2014-00184
- Event Type
- Malfunction
- Date Received
- March 6, 2014
- Date of Event
- February 26, 2014
- Report Date
- February 27, 2014
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K092987
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
IT IS INDICATED THAT PRODUCT IS NOT RETURNING FOR EVALUATION. THEREFORE, INVESTIGATION OF THE COMPLAINT TO DETERMINE ROOT CAUSE CANNOT BE COMPLETED. ROOT CAUSE COULD NOT BE DETERMINED FROM THE INFORMATION PROVIDED BY THE CUSTOMER. SINCE THE PRODUCT ASSOCIATED WITH THE COMPLAINT WAS NOT RETURNED, RETAIN TESTING WAS PERFORMED DURING AN IN-HOUSE INVESTIGATION. RETAIN STRIP TESTING RESULTS MET BOTH ACCURACY AND REPEATABILITY CRITERIA. THE PRODUCT PERFORMED AS EXPECTED AND NO PRODUCT DEFICIENCIES WERE OBSERVED. THE MANUFACTURING RECORDS FOR THE LOT WERE REVIEWED. THE LOT MET SPECIFICATIONS AND NO NON-CONFORMANCES WERE DOCUMENTED. NO FURTHER INVESTIGATION WILL BE PURSUED AT THIS TIME. BASED ON THE INFORMATION AVAILABLE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT INRATIO2 RESULTS. RESULTS AS FOLLOWS: DATE: (B)(6) 2014; INRATIO: 2.9; POC: 1.3. TESTS WERE PERFORMED USING DIFFERENT FINGERS WITHIN 5 MINUTES OF EACH OTHER. THERAPEUTIC RANGE: 2.0-3.0.