ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00435
- Event Type
- Malfunction
- Date Received
- May 26, 2014
- Date of Event
- April 21, 2014
- Report Date
- April 29, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K122432
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- NURSE
Narratives
(B)(4). METHOD: ONLY ONE COMPLAINT RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4) FOR INVESTIGATION. THIS CIRCUIT WAS VISUALLY INSPECTED FOR DAMAGE. RESULTS: A CUT WAS FOUND IN THE EXPIRATORY LIMB OF THE COMPLAINT CIRCUIT. THE CUT WAS ABOUT 68 CM FROM THE PATIENT END. THE DAMAGE APPEARED TO HAVE BEEN MADE WITH A SHARP OBJECT, SUCH AS A KNIFE OR BOX CUTTER. A LOT CHECK HAS REVEALED NO SIMILAR COMPLAINTS FOR LOT NUMBER 131015. CONCLUSION: BASED ON OUR EXPERIENCE OF OTHER SIMILAR INVESTIGATIONS, WE CAN CONCLUDE THAT THE DAMAGE MOST LIKELY OCCURRED WHEN THE CIRCUIT PACKAGING BOX WAS OPENED. ALL RT380 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING ARE PERFORMED EVERY 15 MINUTES. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THE USER INSTRUCTIONS SUPPLIED WITH THE RT380 BREATHING CIRCUIT STATE: - PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. - SET APPROPRIATE VENTILATOR ALARMS. THE HOSPITAL STAFF CORRECTLY CHECKED THE BREATHING CIRCUIT BEFORE PATIENT USE WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE REPRESENTATIVE THAT TWO RT380 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUITS WERE LEAKING. ONE CIRCUIT HAD A VISIBLE CRACK. THIS WAS DISCOVERED BEFORE USE ON A PATIENT WHILST PERFORMING AN AIR TIGHTNESS CHECK ON THE VENTILATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 309979 | ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT380 | 131015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |